RFQ: High-Range Optical Dew Point Hygrometer
National Physical Laboratory Active Tender notice
- Buyer
- National Physical Laboratory
- Procedure
- Competitive quotation (below threshold)
- Category
- Goods
- CPV code
38000000Laboratory, optical and precision equipments (excl. glasses)- First published
- 22 Jul 2024
- Last updated
- 22 Jul 2024
- Rules
- Before the Procurement Act 2023: the Public Contracts Regulations 2015 and earlier rules, or below their thresholds
Description
NPLs requirement is for a high-range hygrometer with the ability to measure up to +95 °C dew-point temperature with an accuracy of ≤ ±0.1 °C. The purpose is to give confidence to the humidity generated by NPL's primary standard generators. Allowing us to perform maintenance activities and ensuring our ability to deliver our stated international CMCs (Calibration & Measurement Capabilities), as well as enabling its use as a reference standard hygrometer for comparison calibrations.
The setup needs to allow for two separate scenarios. In the first circumstance, the test gas will be supplied to the hygrometer inlet from an NPL primary standard generator at 105 kPa and be vented to atmospheric through a rotameter with a needle valve assembly for flow control. In the second circumstance, the hygrometer must have an internal pump that samples gas from a temperature and humidity chamber.
Additional information:
The Authority is using the LUPC Bravo e-Procurement portal to conduct the procurement process. All Suppliers must register their company via the LUPC Bravo Portal.
Notices
- 22 Jul 2024 Tender notice
Contracts Finder holds the official notices, with their documents and contact details.