CEFAS21-01 RFQ- Laboratory Environment Temperature Monitoring Devices
Defra Network eTendering Portal Complete Contract award
- Buyer
- Defra Network eTendering Portal
- Estimated value
- £70,140 as published; Contracts Finder does not say whether VAT is included
- Awarded value
- £70,140 as published, VAT not stated
- Procedure
- Open procedure (below threshold)
- Category
- Services
- CPV code
71700000Monitoring and control services- First published
- 13 Jan 2022
- Last updated
- 13 Jan 2022
- Rules
- Before the Procurement Act 2023: the Public Contracts Regulations 2015 and earlier rules, or below their thresholds
Description
1. Approximately 113 units required for temperature monitoring and compliance suitable for a laboratory environment to monitor temperatures over wide working ranges as depicted below:
Working Temperature range:
Fridges x28 1°C-10°C generally +/- 0.1-10°C (Lowestoft) 4°C +/- 4°C (Weymouth)
Freezers (-40oC) x25 -40oC generally +/- 5°C (Weymouth)
Incubators (including temperature profiling requirements) x33 15-37oC generally +/- 1°C to +/- 2°C tolerance range for individual incubators
Ovens x 1 Max Temperature range: 105oC +/- 10°C (for normal drying ovens - not muffle furnaces)
Minus (-80°C) Freezers x19 Min Temperature range -80°C+/- 8°C
Rooms (ambient) x7 20°C +/- 5°C generally
2. To be able to electronically record this information to comply with our quality standards including ISO9001, ISO17025 and GLP.
3. All items need to be traceable to a reference thermometer either directly or other process.
4. The solution would need to operate within all temperature parameters as stated above in section 1 and include stated uncertainty and offset within range.
5. Full working instructions to be included and consideration of service and maintenance requirements.
6. Requirement to be able to carry out temperature profiling with adjustable time setting for readings and to work over specific ranges for different pieces of equipment.
7. Must have Data logger resolution of 0.02°C, must be accurate to <1°C
8. Out of tolerance alert - email as minimum requirement
9. Potential of calibration also by existing internal methods.
Awards
| Supplier | Value | Date | Status |
|---|---|---|---|
| Henderson Biomedical | £70,140 | 24 Feb 2021 | Active |
Notices
- 13 Jan 2022 Contract award
Contracts Finder holds the official notices, with their documents and contact details.